Subcutaneous ICD and ventricular tachycardia degeneration

Anna Mengoni

Cardiology and cardiovascular physiopathology, S. Maria della Misericordia Hospital, Perugia, Italy

E-mail : bhuvaneswari.bibleraaj@uhsm.nhs.uk

Erberto Carluccio

Cardiology and cardiovascular physiopathology, S. Maria della Misericordia Hospital, Perugia, Italy

Gianluca Zingarini

Cardiology, Maria della Misericordia Hospital, Perugia, Italy

Giuseppe Ambrosio

Cardiology and cardiovascular physiopathology, S. Maria della Misericordia Hospital, Perugia, Italy

Alessandra Tordini

Cardiology and cardiovascular physiopathology, S. Maria della Misericordia Hospital, Perugia, Italy

Paolo Biagioli

Cardiology and cardiovascular physiopathology, S. Maria della Misericordia Hospital, Perugia, Italy

David Giannandrea

UO Neurology, USL Umbria 1 Centro Ictus - Gubbio e Gualdo Tadino Hospital, Italy

Claudio Cavallini

Cardiology, Maria della Misericordia Hospital, Perugia, Italy

DOI: 10.15761/JIC.1000215

Article
Article Info
Author Info
Figures & Data

Abstract

Sudden cardiac death (SCD) remains a leading cause of cardiovascular mortality in the world. The trans-venous implantable cardioverter-defibrillator(TV-ICD) is the most effective approach for primary and secondary prevention of SCD. In response to the growing treatment complexity associated with long-term endocardial management, the subcutaneous implantable cardioverter-defibrillator ICD (SICD) was developed. We report a case in which S-ICD shocks changed a stable ventricular tachycardia (VT) to unstable VT o ventricular fibrillation (VF). 

Case report

We reported the case of a 55 years old man admitted to our hospital because of arrhythmic storm and repeated shocks from a subcutaneous implantable cardioverter-defibrillator ICD (S-ICD). His clinical history was relevant for cardiomyopathy post-myocarditis (2011) with reduced left ventricular (LV) ejection fraction (EF, 40%) due to hypo-akinesia of infero-posterior wall and posterior interventricular septum, without significant coronary artery stenotic lesions at coronary angiography. In the same year, the patient received trans-venous cardioverter-defibrillator (TV-ICD) implant because of non-sustained ventricular tachycardia (VT) episodes during ambulatory ECG Holter monitoring. Subsequently, several appropriate anti-tachycardia pacing interventions (ATP) and shock-therapy occurred. In 2014, the patients developed endocarditis involving the device lead. This leads the device be explanted. Because of the high arrhythmic risk and the impossibility to wait for the end of entire antibiotic therapy, and taking into account the patient’s choice, it was decided to not implant a new TV-ICD in the contralateral side. Therefore, a S- ICD was implanted. The device was programmed to recognize ventricular arrhythmias when 18 out of 24 consecutive sensed events exceed a pre-determined non-programmable detection limit zone; the shock impedances was set to 64 Ohm and the shock energy 80 Joule, with standard shock polarity.

In September  2015 the patient was again admitted to the Hospital because of new arrhythmic storm and repeated shocks from a S-ICD. In the emergency room, he received lidocaine 100 mg iv and amiodarone 300 mg iv, without benefit. At the admission in the Intensive Care Unit, his blood pressure was 100/60 mmHg, heart rate190 beats/min and oxygen saturation 98% in room air. Biochemical laboratory exams did not show any electrolytes imbalance. However, intravenous administration of magnesium sulphate, potassium and esmolol produced a significant attenuation of the arrhythmic events (Figure 1).

Figure 1. Panel A Telemetry: ventricular tachycardia. Panel B Telemetry:Sinus rhythm, ventricular beats, ventricular tachycardia interrupted by a DC-Shock. Panel C Telemetry: Sinus and Ventricular beats. Panel D Electrocardiogram: sinus rhythm, Q wave in DIII and aVF.

The S-ICD interrogation showed several VT episodes, effectively treated by device intervention (Figure 2, panel A). However, after the first S-ICD shocks, several episodes of degeneration of VT in ventricular fibrillation (VF) were also recorded (Figure 2, panel B), which were always interrupted by repeated further S-ICD’s interventions. Therefore, because of repeating shocks, it was first decided to deactivate the S-ICD. Few days later, patients underwent radio frequency ablation of the ventricular arrhythmias, which was not fully therapeutically resolvent. 

Figure 2. S-ICD control. Panel A: Ventricular tachycardia treated effectively with S-ICD shock. Panel B: Ventricular tachycardia degenerated in ventricular fibrillation after S-ICD shock and S-ICD shock not effectively on ventricular fibrillation.

Therefore, the S-ICD was explanted and substituted by a traditional TV-ICD. During the subsequent 12 months follow-up, periodical ambulatory ICD interrogations showed several VT episodes, effectively treated with ATP (Figure 3).

2021 Copyright OAT. All rights reserv

Figure 3. A remote monitoring shows 12 episodes of non sustained ventricular tachycardia and 17 episodes of sustained ventricular tachycardia treated effectively with anti tachycardia pacing.

Discussion

This dramatic case of arrhythmic storm, favourite by a device, underlies the importance to recognize shortcomings of any technology. Review of the literature reports this complication in about 4% of TV-ICD therapy triggered by ATP. However, nowadays no cases have been described for S-ICD has.  Our patient apparently had not precipitant causes of VT (i.e. acute coronary syndrome or electrolytes imbalance); therefore, it is possible that dispersion of ventricular repolarisation and repeated shocks in the vulnerable period could have induced degeneration of VT [1]. 

The dispersion of ventricular repolarisation correlates closely with the vulnerable period to single electrical shocks and several studies have reported a link between the dispersion of repolarisation and the susceptibility to arrhythmias [2-4]. Non-uniform recovery of ventricular excitability plays an important role in the genesis of ventricular arrhythmia; in fact, shock acting in the context of tissue with heterogeneous repolarisation may induce additional action potentials in regions with short repolarisation times and prolonged action potentials in regions with long repolarisation times [5]. Overall, this may facilitate re-entry of multiple activation wave fronts and induction of ventricular fibrillation.

In our patients, the dispersion of ventricular depolarisation is likely to depend on the pathophysiological substrate (fibrosis post-myocarditis).

S-ICD lack of an anti-tachycardia pacing function. 

However, although in our patient several VT episodes were effectively terminated by anti tachycardia pacing, we decided to implant S-ICD before discharge for several reasons: 

  1. As demonstrated by the START trial, both TV-ICD and S-ICD can effectively detect and treat ventricular arrhythmias [6]
  2. many episodes of non-sustained VT, although being treated by ATP, generally are likely to terminate spontaneously [7]
  3. ATP therapy could also be harmful by converting a stable VT to an unstable VT or VF.  Therefore, at the moment, an ideal device does not exist; both TV-ICD and S-ICD are similar for efficacy and safety and the choice may be done only on clinical considerations. 

References

  1. Shibata N, Chen PS, Dixon EG, et al. (1988) Influence of shock strength and timing on induction of ventricular arrhythmias in dogs. Am J Physiol 255:H891-H901. This finding concurs with the hypothesis that the vulnera- bility of the heart relates to the late repolarisation phase of the action potential
  2. Wiggers CJ, Wegren R (1940) Ventricular fibrillation due to single local- ized induction in condenser shock supplied during the vulnerable phase of ventricular systole. Am J Physiol 128:500-505.
  3. Brugada J, Brugada P, Bcersma L, et al. (1991) On the mechanisms of ventricular tachycardia acceleration during programmed electrical stimulation. Circulation 83: 1621-1629.
  4. Taggart P, Sutton P, Lab M, Dean J, Harrison F (1990) Interplay between adrenaline and interbeat interval on ventricular repolarisation in intact heart in vivo. Cardiovasc Res 24:884-895.
  5. Franz MR, Swerdlow CD, Liem LB, Schaefer J (1988) Cycle length dependence of human action potential duration in vivo. Effects of single extrastimuli, sudden sustained rate acceleration and deceleration, and different steady-state frequencies. J Clin Invest 82: 972-979. [Crossref]
  6. Gold MR, Theuns DA, Knight BP, Sturdivant JL, Sanghera R, et al. (2012) Head-to-head comparison of arrhythmia discrimination performance of subcutaneous and transvenous ICD arrhythmia detection algorithms: the START study. J Cardiovasc Electrophysiol 23: 359-366. [Crossref]
  7. Moss AJ, Schuger C, Beck CA, Brown MW, Cannom DS, et al. (2012) Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med 367: 2275-2283. [Crossref]

Editorial Information

Editor-in-Chief

Massimo Fioranelli
Guglielmo Marconi University

Article Type

Case Report

Publication history

Received date: March 12, 2017
Accepted date: April 04, 2017
Published date: April 07, 2017

Copyright

© 2017 Anna Mengoni. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Citation

Mengoni A, Carluccio E, Zingarini G, AmbrosioG, Tordini T1, et al. (2016) Subcutaneous ICD and ventricular tachycardia degeneration. J Integr Cardiol 3: DOI: 10.15761/JIC.1000215

Corresponding author

Anna Mengoni

Cardiology and cardiovascular physiopathology, S Maria della Misericordia Hospital, Perugia, Italy; Tel: 0755782394

Figure 1. Panel A Telemetry: ventricular tachycardia. Panel B Telemetry:Sinus rhythm, ventricular beats, ventricular tachycardia interrupted by a DC-Shock. Panel C Telemetry: Sinus and Ventricular beats. Panel D Electrocardiogram: sinus rhythm, Q wave in DIII and aVF.

Figure 2. S-ICD control. Panel A: Ventricular tachycardia treated effectively with S-ICD shock. Panel B: Ventricular tachycardia degenerated in ventricular fibrillation after S-ICD shock and S-ICD shock not effectively on ventricular fibrillation.

Figure 3. A remote monitoring shows 12 episodes of non sustained ventricular tachycardia and 17 episodes of sustained ventricular tachycardia treated effectively with anti tachycardia pacing.